Nalaganje...
The Fabrication and Characterization of Ni/4H-SiC Schottky Diode Radiation Detectors with a Sensitive Area of up to 4 cm(2)
Silicon carbide (SiC) detectors of an Ni/4H-SiC Schottky diode structure and with sensitive areas of 1–4 cm(2) were fabricated using high-quality lightly doped epitaxial 4H-SiC material, and were tested in the detection of alpha particles and pulsed X-rays/UV-light. A linear energy response to alpha...
Shranjeno v:
| izdano v: | Sensors (Basel) |
|---|---|
| Main Authors: | , , , , , , , , , , |
| Format: | Artigo |
| Jezik: | Inglês |
| Izdano: |
MDPI
2017
|
| Teme: | |
| Online dostop: | https://ncbi.nlm.nih.gov/pmc/articles/PMC5677359/ https://ncbi.nlm.nih.gov/pubmed/29027944 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.3390/s17102334 |
| Oznake: |
Označite
Brez oznak, prvi označite!
|