Deceleration of probe beam by stage bias potential improves resolution of serial block-face scanning electron microscopic images
Serial block-face scanning electron microscopy (SBEM) is quickly becoming an important imaging tool to explore three-dimensional biological structure across spatial scales. At probe-beam-electron energies of 2.0 keV or lower, the axial resolution should improve, because there is less primary electro...