An integrated Taguchi and response surface methodological approach for the optimization of an HPLC method to determine glimepiride in a supersaturatable self-nanoemulsifying formulation
We studied the application of Taguchi orthogonal array (TOA) design during the development of an isocratic stability indicating HPLC method for glimepiride as per TOA design; twenty-seven experiments were conducted by varying six chromatographic factors. Percentage of organic phase was the most sign...