Ładuje się......
Energy and coincidence time resolution measurements of CdTe detectors for PET
We report on the characterization of 2 mm thick CdTe diode detector with Schottky contacts to be employed in a novel conceptual design of PET scanner. Results at −8°C with an applied bias voltage of −1000 V/mm show a 1.2% FWHM energy resolution at 511 keV. Coincidence time resolution has been measur...
Zapisane w:
| Główni autorzy: | , , , , , , , , , , |
|---|---|
| Format: | Artigo |
| Język: | Inglês |
| Wydane: |
2013
|
| Hasła przedmiotowe: | |
| Dostęp online: | https://ncbi.nlm.nih.gov/pmc/articles/PMC3672966/ https://ncbi.nlm.nih.gov/pubmed/23750177 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1088/1748-0221/8/02/C02015 |
| Etykiety: |
Dodaj etykietę
Nie ma etykietki, Dołącz pierwszą etykiete!
|