A carregar...

Energy and coincidence time resolution measurements of CdTe detectors for PET

We report on the characterization of 2 mm thick CdTe diode detector with Schottky contacts to be employed in a novel conceptual design of PET scanner. Results at −8°C with an applied bias voltage of −1000 V/mm show a 1.2% FWHM energy resolution at 511 keV. Coincidence time resolution has been measur...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Main Authors: Ariño, G., Chmeissani, M., De Lorenzo, G., Puigdengoles, C., Cabruja, E., Calderón, Y., Kolstein, M., Macias-Montero, J.G., Martinez, R., Mikhaylova, E., Uzun, D.
Formato: Artigo
Idioma:Inglês
Publicado em: 2013
Assuntos:
Acesso em linha:https://ncbi.nlm.nih.gov/pmc/articles/PMC3672966/
https://ncbi.nlm.nih.gov/pubmed/23750177
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1088/1748-0221/8/02/C02015
Tags: Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!