A carregar...
Energy and coincidence time resolution measurements of CdTe detectors for PET
We report on the characterization of 2 mm thick CdTe diode detector with Schottky contacts to be employed in a novel conceptual design of PET scanner. Results at −8°C with an applied bias voltage of −1000 V/mm show a 1.2% FWHM energy resolution at 511 keV. Coincidence time resolution has been measur...
Na minha lista:
| Main Authors: | , , , , , , , , , , |
|---|---|
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
2013
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC3672966/ https://ncbi.nlm.nih.gov/pubmed/23750177 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1088/1748-0221/8/02/C02015 |
| Tags: |
Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
|