Direct detection and identification of active pharmaceutical ingredients in intact tablets by helium plasma ionization (HePI) mass spectrometry
A simple modification converts an electrospray ion source to an ambient-pressure helium plasma ionization source without the need of additional expensive hardware. Peaks for active ingredients were observed in the spectra recorded from intact pharmaceutical tablets placed in this source. A flow of h...