Dosimetric and mechanical equivalency of Varian TrueBeam linear accelerators
PURPOSE: To investigate and improve the level of equivalency of Varian TrueBeam linear accelerators (linacs) in energy‐, dosimetric leaf gap‐ (DLG) and jaw calibration. METHODS: Eight linacs with four photon energies: 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV, and three electron energies (on two linacs):...