Nalaganje...
Dosimetric and mechanical equivalency of Varian TrueBeam linear accelerators
PURPOSE: To investigate and improve the level of equivalency of Varian TrueBeam linear accelerators (linacs) in energy‐, dosimetric leaf gap‐ (DLG) and jaw calibration. METHODS: Eight linacs with four photon energies: 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV, and three electron energies (on two linacs):...
Shranjeno v:
| izdano v: | J Appl Clin Med Phys |
|---|---|
| Main Authors: | , , , , |
| Format: | Artigo |
| Jezik: | Inglês |
| Izdano: |
John Wiley and Sons Inc.
2020
|
| Teme: | |
| Online dostop: | https://ncbi.nlm.nih.gov/pmc/articles/PMC7769408/ https://ncbi.nlm.nih.gov/pubmed/33070456 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/acm2.13058 |
| Oznake: |
Označite
Brez oznak, prvi označite!
|