Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up

A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD...