A carregar...

Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up

A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Publicado no:J Arrhythm
Main Authors: Hori, Yuichi, Nakahara, Shiro, Nishiyama, Naoki, Fukuda, Reiko, Toratani, Noritaka, Sakai, Yoshihiko, Taguchi, Isao
Formato: Artigo
Idioma:Inglês
Publicado em: John Wiley and Sons Inc. 2018
Assuntos:
Acesso em linha:https://ncbi.nlm.nih.gov/pmc/articles/PMC6174498/
https://ncbi.nlm.nih.gov/pubmed/30327706
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/joa3.12104
Tags: Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!