Normal tissue doses from MV image‐guided radiation therapy (IGRT) using orthogonal MV and MV‐CBCT
PURPOSE: The aim of this study was to measure and compare the mega‐voltage imaging dose from the Halcyon medical linear accelerator (Varian Medical Systems) with measured imaging doses with the dose calculated by Eclipse treatment planning system. METHODS: An anthropomorphic thorax phantom was image...