Independent validation of machine performance check for the Halcyon and TrueBeam linacs for daily quality assurance
PURPOSE: To evaluate the ability of the machine performance check (MPC) on the Halcyon to detect errors, with comparison with the TrueBeam. METHODS: MPC is an automated set of quality assurance (QA) tests that use a phantom placed on the couch and the linac's imaging system(s) to verify the bea...