Beam focal spot position: The forgotten linac QA parameter. An EPID‐based phantomless method for routine Stereotactic linac QA

Modern day Stereotactic treatments require high geometric accuracy of the delivered treatment. To achieve the required accuracy the IGRT imaging isocenter needs to closely coincide with the treatment beam isocenter. An influence on this isocenter coincidence and on the spatial positioning of the bea...