A carregar...

Beam focal spot position: The forgotten linac QA parameter. An EPID‐based phantomless method for routine Stereotactic linac QA

Modern day Stereotactic treatments require high geometric accuracy of the delivered treatment. To achieve the required accuracy the IGRT imaging isocenter needs to closely coincide with the treatment beam isocenter. An influence on this isocenter coincidence and on the spatial positioning of the bea...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Publicado no:J Appl Clin Med Phys
Main Authors: Chojnowski, Jacek M., Barnes, Michael P., Sykes, Jonathan R., Thwaites, David I.
Formato: Artigo
Idioma:Inglês
Publicado em: John Wiley and Sons Inc. 2017
Assuntos:
Acesso em linha:https://ncbi.nlm.nih.gov/pmc/articles/PMC5875833/
https://ncbi.nlm.nih.gov/pubmed/28786168
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/acm2.12147
Tags: Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!