A carregar...
Angular dependence of the output of a kilovoltage X‐ray therapy unit
During the recommissioning of a Philips RT‐250 kilovoltage X‐ray unit, unexpected output variations with tube head rotation (cross‐plane) and tube head tilt (in‐plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0° in‐plane and 0° cross‐plane...
Na minha lista:
| Publicado no: | J Appl Clin Med Phys |
|---|---|
| Main Authors: | , , , |
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
John Wiley and Sons Inc.
2010
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC5720435/ https://ncbi.nlm.nih.gov/pubmed/20717089 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1120/jacmp.v11i3.3216 |
| Tags: |
Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
|