MRI Compatibility of Microfabricated Magnetic Actuators for Implantable Catheters: Mechanical Evaluations

Here we demonstrate the mechanical robustness of microfabricated torsional magnetic actuators in withstanding the magnetic fields produced by a 7 T MRI magnet. The static and dynamic mechanical characteristics of 30 devices were quantitatively measured before and after exposure to both uniform and n...