Lataa...

A preclinical numerical assessment of a polyetheretherketone femoral component in total knee arthroplasty during gait

BACKGROUND: Conventional total knee replacement designs show high success rates but in the long term, the stiff metal components may affect bone quality of the distal femur. In this study we introduce an all-polymer total knee replacement device containing a PEEK femoral component on an UHMWPE tibia...

Täydet tiedot

Tallennettuna:
Bibliografiset tiedot
Julkaisussa:J Exp Orthop
Päätekijät: de Ruiter, Lennert, Janssen, Dennis, Briscoe, Adam, Verdonschot, Nico
Aineistotyyppi: Artigo
Kieli:Inglês
Julkaistu: Springer Berlin Heidelberg 2017
Aiheet:
Linkit:https://ncbi.nlm.nih.gov/pmc/articles/PMC5285288/
https://ncbi.nlm.nih.gov/pubmed/28144820
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1186/s40634-017-0078-4
Tagit: Lisää tagi
Ei tageja, Lisää ensimmäinen tagi!