Monte Carlo Study of the Effect of Collimator Thickness on T-99m Source Response in Single Photon Emission Computed Tomography

In single photon emission computed tomography (SPECT), the collimator is a crucial element of the imaging chain and controls the noise resolution tradeoff of the collected data. The current study is an evaluation of the effects of different thicknesses of a low-energy high-resolution (LEHR) collimat...