Carregant...

Monte Carlo Study of the Effect of Collimator Thickness on T-99m Source Response in Single Photon Emission Computed Tomography

In single photon emission computed tomography (SPECT), the collimator is a crucial element of the imaging chain and controls the noise resolution tradeoff of the collected data. The current study is an evaluation of the effects of different thicknesses of a low-energy high-resolution (LEHR) collimat...

Descripció completa

Guardat en:
Dades bibliogràfiques
Autors principals: Islamian, Jalil Pirayesh, Toossi, Mohammad Taghi Bahreyni, Momennezhad, Mahdi, Zakavi, Seyyed Rasoul, Sadeghi, Ramin, Ljungberg, Michael
Format: Artigo
Idioma:Inglês
Publicat: Medknow Publications & Media Pvt Ltd 2012
Matèries:
Accés en línia:https://ncbi.nlm.nih.gov/pmc/articles/PMC3555397/
https://ncbi.nlm.nih.gov/pubmed/23372440
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.4103/1450-1147.103419
Etiquetes: Afegir etiqueta
Sense etiquetes, Sigues el primer a etiquetar aquest registre!