A phantom study of an in vivo dosimetry system using plastic scintillation detectors for real-time verification of (192)Ir HDR brachytherapy

Purpose: The goal of the present work was to evaluate the accuracy of a plastic scintillation detector (PSD) system to perform in-phantom dosimetry during (192)Ir high dose rate (HDR) brachytherapy treatments. Methods: A PSD system capable of stem effect removal was built. A red–green–blue photodiod...