Lataa...

A phantom study of an in vivo dosimetry system using plastic scintillation detectors for real-time verification of (192)Ir HDR brachytherapy

Purpose: The goal of the present work was to evaluate the accuracy of a plastic scintillation detector (PSD) system to perform in-phantom dosimetry during (192)Ir high dose rate (HDR) brachytherapy treatments. Methods: A PSD system capable of stem effect removal was built. A red–green–blue photodiod...

Täydet tiedot

Tallennettuna:
Bibliografiset tiedot
Päätekijät: Therriault-Proulx, Francois, Briere, Tina M., Mourtada, Firas, Aubin, Sylviane, Beddar, Sam, Beaulieu, Luc
Aineistotyyppi: Artigo
Kieli:Inglês
Julkaistu: American Association of Physicists in Medicine 2011
Aiheet:
Linkit:https://ncbi.nlm.nih.gov/pmc/articles/PMC3104721/
https://ncbi.nlm.nih.gov/pubmed/21776789
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1118/1.3572229
Tagit: Lisää tagi
Ei tageja, Lisää ensimmäinen tagi!