Verification of tomotherapy dose delivery
Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% a...