A carregar...
Verification of tomotherapy dose delivery
Seventy-one patient-specific delivery quality assurance (DQA) plans for the Tomotherapy HI-ART II helical tomotherapy system (TomoTherapy, Inc., Madison, WI, USA) were measured using film and ion chamber. The agreement in absolute point dose was 1.19 ± 0.79%, 1.91 ± 1.39%, 2.14 ± 1.3%, 1.3 ± 0.73% a...
Na minha lista:
| Main Authors: | , |
|---|---|
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
Medknow Publications
2009
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC2807687/ https://ncbi.nlm.nih.gov/pubmed/20098569 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.4103/0971-6203.54856 |
| Tags: |
Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
|