Lanean...

Influence of conductance changes on patch clamp capacitance measurements using a lock-in amplifier and limitations of the phase tracking technique.

We characterized the influence of conductance changes on whole-cell patch clamp capacitance measurements with a lock-in amplifier and the limitations of the phase-tracking method by numerical computer simulations, error formulas, and experimental tests. At correct phase setting, the artifacts in the...

Deskribapen osoa

Gorde:
Xehetasun bibliografikoak
Egile Nagusiak: Debus, K, Hartmann, J, Kilic, G, Lindau, M
Formatua: Artigo
Hizkuntza:Inglês
Argitaratua: 1995
Gaiak:
Sarrera elektronikoa:https://ncbi.nlm.nih.gov/pmc/articles/PMC1236518/
https://ncbi.nlm.nih.gov/pubmed/8599687