Lanean...
Influence of conductance changes on patch clamp capacitance measurements using a lock-in amplifier and limitations of the phase tracking technique.
We characterized the influence of conductance changes on whole-cell patch clamp capacitance measurements with a lock-in amplifier and the limitations of the phase-tracking method by numerical computer simulations, error formulas, and experimental tests. At correct phase setting, the artifacts in the...
Gorde:
| Egile Nagusiak: | , , , |
|---|---|
| Formatua: | Artigo |
| Hizkuntza: | Inglês |
| Argitaratua: |
1995
|
| Gaiak: | |
| Sarrera elektronikoa: | https://ncbi.nlm.nih.gov/pmc/articles/PMC1236518/ https://ncbi.nlm.nih.gov/pubmed/8599687 |
| Etiketak: |
Etiketa erantsi