Optimization and validation of an RP-HPLC method for the estimation of 6-mercaptopurine in bulk and pharmaceutical formulations
A reverse phase HPLC method is described for the determination of 6-mercaptopurine in bulk and tablets. Chromatography was carried on a C18 column using a mixture of acetonitrile and 0.05 mol/L sodium acetate buffer (10:90 v/v) as the mobile phase at a flow rate of 1 mL/min-1 with detection at 324 n...