Optimization and validation of an RP-HPLC method for the estimation of 6-mercaptopurine in bulk and pharmaceutical formulations
A reverse phase HPLC method is described for the determination of 6-mercaptopurine in bulk and tablets. Chromatography was carried on a C18 column using a mixture of acetonitrile and 0.05 mol/L sodium acetate buffer (10:90 v/v) as the mobile phase at a flow rate of 1 mL/min-1 with detection at 324 n...
Na minha lista:
| Autor principal: | |
|---|---|
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
Universidade de São Paulo, Faculdade de Ciências Farmacêuticas
2014-12-01
|
| coleção: | Brazilian Journal of Pharmaceutical Sciences |
| Assuntos: | |
| Acesso em linha: | http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1984-82502014000400793&lng=en&tlng=en |
| Tags: |
Sem tags, seja o primeiro a adicionar uma tag!
|
