A carregar...

MR‐guided proton therapy: Impact of magnetic fields on the detector response

PURPOSE: To investigate the response of detectors for proton dosimetry in the presence of magnetic fields. MATERIAL AND METHODS: Four ionization chambers (ICs), two thimble‐type and two plane‐parallel‐type, and a diamond detector were investigated. All detectors were irradiated with homogeneous sing...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Publicado no:Med Phys
Main Authors: Fuchs, Hermann, Padilla‐Cabal, Fatima, Zimmermann, Lukas, Palmans, Hugo, Georg, Dietmar
Formato: Artigo
Idioma:Inglês
Publicado em: John Wiley and Sons Inc. 2021
Assuntos:
Acesso em linha:https://ncbi.nlm.nih.gov/pmc/articles/PMC8251909/