The reliability of radiation dose display of a computed tomography scanner

BACKGROUND: Internationally, the typical allowed difference between the measured radiation dose and dose reported by a computed tomography (CT) scanner is ±20 %. The objective is to describe a method in order to analyse this difference in a CT scanner in the Emergency Department of Kanta-Häme Centra...