Monte Carlo simulation of 6‐MV dynamic wave VMAT deliveries by Vero4DRT linear accelerator using EGSnrc moving sources
The commissioning and benchmark of a Monte Carlo (MC) model of the 6‐MV Brainlab‐Mitsubishi Vero4DRT linear accelerator for the purpose of quality assurance of clinical dynamic wave arc (DWA) treatment plans is reported. Open‐source MC applications based on EGSnrc particle transport codes are used t...