Reliable phase-contrast flow volume magnetic resonance measurements are feasible without adjustment of the velocity encoding parameter

Purpose: To show that adjustment of velocity encoding (VENC) for phase-contrast (PC) flow volume measurements is not necessary in modern MR scanners with effective background velocity offset corrections. Approach: The independence on VENC was demonstrated theoretically, but also experimentally on de...