Small field output correction factors of the microSilicon detector and a deeper understanding of their origin by quantifying perturbation factors

PURPOSE: The aim of this study is the experimental and Monte Carlo‐based determination of small field correction factors for the unshielded silicon detector microSilicon for a standard linear accelerator as well as the Cyberknife System. In addition, a detailed Monte Carlo analysis has been performe...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Publicado no:Med Phys
Main Authors: Weber, Carolin, Kranzer, Rafael, Weidner, Jan, Kröninger, Kevin, Poppe, Björn, Looe, Hui Khee, Poppinga, Daniela
Formato: Artigo
Idioma:Inglês
Publicado em: John Wiley and Sons Inc. 2020
Assuntos:
Acesso em linha:https://ncbi.nlm.nih.gov/pmc/articles/PMC7496769/
https://ncbi.nlm.nih.gov/pubmed/32196683
https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/mp.14149