False elevation of serum cortisol in chemiluminescence immunoassay by Siemens Advia Centaur XP system in 21-hydroxylase deficiency: an ‘endocrine laboma’

Liquid chromatography–mass spectrometry, the gold standard method for cortisol measurement, is expensive and not widely available in the developing countries. Chemiluminescent immunoassay, commonly used for cortisol measurement is prone to clinically meaningful inter-assay variability in some analys...