Cargando...

A 1.5-T Magic-Angle Spinning NMR Magnet: 4.2-K Performance and Field Mapping Test Results

We present results of full-current testing at 4.2 K of a z-axis 0.866-T solenoid and an x-axis 1.225-T dipole coil that comprise a 1.5-T/75-mm room temperature bore magic-angle-spinning nuclear magnetic resonance magnet developed at the MIT Francis Bitter Magnet Laboratory. Also included in the pape...

Descripción completa

Guardado en:
Detalles Bibliográficos
Publicado en:IEEE Trans Appl Supercond
Autores principales: Voccio, John, Hahn, Seungyong, Kim, Youngjae, Song, Jungbin, Kajikawa, Kazuhiro, Noguchi, So, Bascuñán, Juan, Iwasa, Yukikazu