Biomechanical Comparison of a New Memory Compression Alloy Plate versus Traditional Titanium Plate for Anterior Cervical Discectomy and Fusion: A Finite Element Analysis

OBJECTIVE: To compare the biomechanical properties of a new memory compression alloy plate and traditional titanium plate after anterior cervical discectomy and fusion (ACDF). METHODS: A finite element model of the C3-7 segments was developed and validated. The C5-6 disc was removed, and an interver...