Twiddler's Syndrome: An Unusual Cause of Repeated Shocks by Implantable Cardioverter-Defibrillator in an Asymptomatic Patient

Twiddler's syndrome refers to malfunction of pacemaker or implantable cardioverter-defibrillator (ICD) resulting from deliberate or unintentional twisting of pulse generator in device pocket, resulting in lead dislodgment. We report a case of a 65-year-old woman with multiple ICD shocks and ant...