Peripheral Nerve Stimulation Limits of a High Amplitude and Slew Rate Magnetic Field Gradient Coil for Neuroimaging
PURPOSE: To establish peripheral nerve stimulation (PNS) thresholds for an ultra-high performance magnetic field gradient sub-system (simultaneous 200 mT/m gradient amplitude and 500 T/m/s gradient slew rate; 1-MVA per-axis: referred to as MAGNUS) designed for neuroimaging with asymmetric transverse...
Guardat en:
| Publicat a: | Magn Reson Med |
|---|---|
| Autors principals: | , , , , , , , |
| Format: | Artigo |
| Idioma: | Inglês |
| Publicat: |
2019
|
| Matèries: | |
| Accés en línia: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6778706/ https://ncbi.nlm.nih.gov/pubmed/31385628 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/mrm.27909 |
| Etiquetes: |
Afegir etiqueta
Sense etiquetes, Sigues el primer a etiquetar aquest registre!
|