The Graphite Calorimeter as a Standard of Absorbed Dose for Cobalt-60 Gamma Radiation
Absorbed dose to water in a cobalt-60 gamma-ray beam has been determined using a thick-walled graphite ionization chamber. The chamber was calibrated in a graphite phantom against a graphite calorimeter, and the graphite calibration factor was converted to a water calibration factor using published...
Сохранить в:
| Опубликовано в: : | J Res Natl Bur Stand (1977) |
|---|---|
| Главные авторы: | , , |
| Формат: | Artigo |
| Язык: | Inglês |
| Опубликовано: |
National Institute of Standards and Technology
1981
|
| Предметы: | |
| Online-ссылка: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6753002/ https://ncbi.nlm.nih.gov/pubmed/34566056 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.6028/jres.086.021 |