Caricamento...
The International Practical Temperature Scale of 1968 in the Region 13.81 K to 90.188 K as Maintained at the National Bureau of Standards
The reproducibility of the International Practical Temperature Scale of 1968 (IPTS–68) in the region of 13.81 K to 90.188 K using standard platinum resistance thermometers is discussed. The effect of errors above 0 °C on the calibration between 13 and 90 K is analyzed. Calibrations resulting from le...
Salvato in:
| Pubblicato in: | J Res Natl Bur Stand A Phys Chem |
|---|---|
| Autori principali: | , , |
| Natura: | Artigo |
| Lingua: | Inglês |
| Pubblicazione: |
National Institute of Standards and Technology
1973
|
| Soggetti: | |
| Accesso online: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6715979/ https://ncbi.nlm.nih.gov/pubmed/32189741 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.6028/jres.077A.019 |
| Tags: |
Aggiungi Tag
Nessun Tag, puoi essere il primo ad aggiungerne! !
|