Helical blade compression failure occurred during PFNA implantation: A rare case and ingenious solution
RATIONALE: Femoral intertrochanteric fracture happens easily in the elderly, especially those with osteoporosis. As a standard intramedullary fixation implant, Proximal Femoral Nail Anti-rotation (PFNA) is applied to various types of stable or unstable intertrochanteric fractures of femur. Due to bl...