Loading...
Determination of hemolysis index thresholds for biochemical tests on Siemens Advia 2400 chemistry analyzer
BACKGROUND: In vitro hemolysis is still the most common source of pre‐analytical nonconformities. This study aimed to investigate the hemolytic effects on commonly used biochemical tests as well as to determine the hemolysis index (HI) thresholds on Siemens Advia 2400 chemistry analyzer. METHODS: Pe...
Na minha lista:
| Udgivet i: | J Clin Lab Anal |
|---|---|
| Main Authors: | , , , , , |
| Format: | Artigo |
| Sprog: | Inglês |
| Udgivet: |
John Wiley and Sons Inc.
2019
|
| Fag: | |
| Online adgang: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6589729/ https://ncbi.nlm.nih.gov/pubmed/30779463 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/jcla.22856 |
| Tags: |
Tilføj Tag
Ingen Tags, Vær først til at tagge denne postø!
|