Evaluation of the Bio-Rad Geenius HIV 1/2 Assay as Part of a Confirmatory HIV Testing Strategy for Quebec, Canada: Comparison with Western Blot and Inno-Lia Assays
The rapid confirmatory Bio-Rad Geenius HIV 1/2 assay was evaluated as an alternative to the HIV-1 Western blot (WB) confirmatory assay. A total of 370 retrospective samples collected from 356 patients were tested. Sensitivity of the Geenius assay to detect HIV-1 and HIV-2 infections was 100% and 97%...