A novel method for monitoring the constancy of beam path accuracy in CyberKnife
The aim of current work was to present a novel evaluation procedure implemented for checking the constancy of beam path accuracy of a CyberKnife system based on ArcCHECK. A tailor‐made Styrofoam with four implanted fiducial markers was adopted to enable the fiducial tracking during beam deliveries....