Comparative Accuracy Evaluation of a Blood Glucose Meter With Novel Hematocrit Correction Technology, With Three Currently Used Commercially Available Blood Glucose Monitoring Systems
Hematocrit is known to influence glucose values obtained on some blood glucose meters, with bias observed especially at low and high hematocrit levels. We evaluated the performance of a meter with hematocrit correction technology alongside 3 other commercially available meters. Capillary blood sampl...
Na minha lista:
| Publicado no: | J Diabetes Sci Technol |
|---|---|
| Main Authors: | , , |
| Formato: | Artigo |
| Idioma: | Inglês |
| Publicado em: |
SAGE Publications
2019
|
| Assuntos: | |
| Acesso em linha: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6501526/ https://ncbi.nlm.nih.gov/pubmed/30623673 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1177/1932296818821389 |