Načítá se...
Edge Packing Artifact Mimicking Lytic–Sclerotic Lesion in Bone Scan Due to Spontaneous Malfunction of Photomultiplier Tubes in the Gamma Camera
Edge packing is a kind of nonuniformity artifact, which usually appears as an accumulation of counts at the edge of the field of view compared to the central region. The common causes of edge packing artifact in the image are defective collimator, noncalibrated photomultiplier tube (PMT), and improp...
Uloženo v:
| Vydáno v: | Indian J Nucl Med |
|---|---|
| Hlavní autoři: | , , , , |
| Médium: | Artigo |
| Jazyk: | Inglês |
| Vydáno: |
Wolters Kluwer - Medknow
2019
|
| Témata: | |
| On-line přístup: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6481219/ https://ncbi.nlm.nih.gov/pubmed/31040526 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.4103/ijnm.IJNM_168_18 |
| Tagy: |
Přidat tag
Žádné tagy, Buďte první, kdo otaguje tento záznam!
|