Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom

PURPOSE: We sought to examine the possibility of reducing the contrast medium dosage in dual-energy imaging using a saline-mixed injection with a virtual monochromatic energy method of dual-source computed tomography (CT). METHODS: An X-ray CT (SOMATOM Definition Flash: Siemens, Nurnberg, Germany) w...