A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process control
BACKGROUND: The automated and integrated machine performance check (MPC) tool was verified against independent detectors to evaluate its beam uniformity and output detection abilities to consider it suitable for daily quality assurance (QA). METHODS: Measurements were carried out on six linear accel...