Ładuje się......
EPID‐based in vivo dosimetry using Dosimetry Check™: Overview and clinical experience in a 5‐yr study including breast, lung, prostate, and head and neck cancer patients
BACKGROUND: Independent verification of the dose delivered by complex radiotherapy can be performed by electronic portal imaging device (EPID) dosimetry. This paper presents 5‐yr EPID in vivo dosimetry (IVD) data obtained using the Dosimetry Check (DC) software on a large cohort including breast, lu...
Zapisane w:
| Wydane w: | J Appl Clin Med Phys |
|---|---|
| Główni autorzy: | , , , , , , , , , , , |
| Format: | Artigo |
| Język: | Inglês |
| Wydane: |
John Wiley and Sons Inc.
2018
|
| Hasła przedmiotowe: | |
| Dostęp online: | https://ncbi.nlm.nih.gov/pmc/articles/PMC6333145/ https://ncbi.nlm.nih.gov/pubmed/30536528 https://ncbi.nlm.nih.govhttp://dx.doi.org/10.1002/acm2.12441 |
| Etykiety: |
Dodaj etykietę
Nie ma etykietki, Dołącz pierwszą etykiete!
|