EPID‐based in vivo dosimetry using Dosimetry Check™: Overview and clinical experience in a 5‐yr study including breast, lung, prostate, and head and neck cancer patients

BACKGROUND: Independent verification of the dose delivered by complex radiotherapy can be performed by electronic portal imaging device (EPID) dosimetry. This paper presents 5‐yr EPID in vivo dosimetry (IVD) data obtained using the Dosimetry Check (DC) software on a large cohort including breast, lu...

Szczegółowa specyfikacja

Zapisane w:
Opis bibliograficzny
Wydane w:J Appl Clin Med Phys
Główni autorzy: Nailon, William H., Welsh, Daniel, McDonald, Kim, Burns, Donna, Forsyth, Julie, Cooke, Gillian, Cutanda, Francisco, Carruthers, Linda J., McLaren, Duncan B., Puxeu Vaqué, Josep, Kehoe, Terence, Andiappa, Sankar