Transient hyperkalemia-induced T-wave oversensing as a cause of inappropriate ICD discharges

The present case describes a patient with an implantable cardioverter-defibrillator (ICD) who reported several ICD inappropriate discharges due to transient T-wave oversensing in the setting of acute hyperkalemia. T-wave oversensing was detected only at telemetry interrogation while T-wave amplitude...