Measuring PET Spatial Resolution Using a Cylinder Phantom Positioned at an Oblique Angle
A cylinder phantom positioned at a slightly oblique angle with respect to the z-axis of a PET scanner allows for fine sampling of the edge-spread function. We show how this technique can be used to measure the spatial resolution that can be expected with clinical PET protocols, potentially providing...