Significance of Daily Quality Assurance Scan in Hardware Artifact Evaluation
A failure in any of the hardware components can lead to degradation in image quality and accuracy of quantification. Few artifacts may bring serious impacts on the image quality and finally lead to the wrong diagnosis. We encountered a sudden appearance of the cold area, being more prominently visib...